Biosafety Cabinet(BSC-01/02ⅡA2)

Friendly Control Panel

  • LED screen, show the air velocity, time, fan station, alarm in real time.

Safe and Reliable, Energy-Efficient

  • DC fan and LED light make the operation power consumption save 30%.
  • Intelligent auto-supply system of imported blower ensures the change of air volume less than 10% when the resistance of filter increases 50% and enhance safety.
  • The alarm system for safe operation height is interlocked with light. At the same time light is Interlocked with sterilizing.
  • Leak detection in cabinet, no leakage at 500Pa.
  • UV light with set timer, safer and high efficiency.

Humanized design

  • 10°sloped front make it comfortable to operator.
  • Negative pressure and double layers chamber around ensure non-leakage, non-pollution. Demountable circular-arc armrest can alleviate working tiredness and make it easy to transport.
  • Front window adopts hanging lifting system (Imported from Japan) and more than 5mm thickness security glass with arbitrary positioning, high stability and convenient to sterilization when it closed.
  • Movable stainless steel worktable 4 liter capacity stainless steel collecting tank with below down valve, easy to collect water, cleaning and disinfection.
  • Non-obstructive air return mode in the front suction channel.
  • The adjustable leveling foot is without exposure thread.

HEPA filters

  • USA HV HEPA Filter Medium, provide 99.995% typical efficiency for particle size over 0.3 microns, and the materials is glass fiber.
  • A minimum ISO 5 air quality in the work area
  • Unique Technology of non leakage.
  • Every HEPA filters will be tested to ensure non leakage.

Specifications

Model/Item

BSC-1301ⅡA2

BSC-1302ⅡA2

Nominal Size 1.3meter(4')

Filtration Efficiency

HEPA: ≥99.995%, @0.3μm/ULPA:≥99.9995%, @0.12μm

Downflow Velocity

0.35 m/s(68.88fpm)

Inflow Velocity

0.54 m/s(106.27fpm)

Noise

≤67 dB(A)

Vibration(μm) ≤5 μm(center of tabletop)

Power Supply

AC115V,1φ,60Hz

AC230V,1φ,50/60Hz

Opening Height(mm) 250

Operator Protection

A. Total colony in impaction sampler ≤10CFU./time

B. Total colony in slot sampler ≤5CFU./time

Product Protection

Total colony in culture dish ≤5CFU./time

Cross Contamination

Total colony in culture dish ≤2CFU./time

Max Consumption(with spare socket)

1.5 kVA

/

Rated Power(without spare socket) 1 kVA 0.46 kVA

Weight

300 kg

 

Work Dimension

(W1XD1XH1) 

 

1270*600*620 mm

(50''x23.6''x24.4'')

 

Overall Dimension

(WXDXH) 

 

1500*790*2050 mm

(59.1''x31.1''x80.7'')

Size and Qty. of

Light

LED 24.5 w x ②

Size and Qty. of

UV Light

15 w x ②

Lumin. 

≥750 Lx

Cabinet Materials High-grade steel and lacquered in ivory
Working Area Materials SS304 fully finished

Air Direction

Top out

Plug Style

USA Type(AC125V/10A)

/

Qty. of Spare Socket

USA Type(AC125V/2A) 2 units

/